Cabinet temperature is 10°F (or 5°C) higher than set point. Silence the alarm by pressing
ALARM RESET (Up â–² arrow) and open door(s) to allow the chamber(s) to cool.
Input temperature is lower than input range. Check temperature probe and connections.
See below for additional troubleshooting.
Input temperature is higher than input range. Check temperature probe and
connections. See below for additional troubleshooting.
Temperature probe is at fault. Check connection, then replace and calibrate.
Control failure. Check connection first, then replace and calibrate.
1. Check outlet for power
2. Check if warmer is plugged in
3. Check if the ON/OFF switch/circuit breaker is turned on
4. Check the fuse on the incoming supply
5. Check for power at the junction box in the control panel
6. Contact your Logiquip authorized service representative
1. Is the power turned on?
2. Is the temperature set above chamber ambient temperature?
3. Is the circulation fan operational? Open the door and press in on the door switch
4. Is there voltage on the output terminal of the controller?
5. Is the door(s) closed?
6. Contact your Logiquip authorized service representative
1. Is the circulating fan operational?
2. Are the contents loaded properly?
3. Has the chamber set temperature been lowered?
4. Temperature of the lower chamber cannot be in excess of +30°F (+1.1°C) above the
upper chamber
5. Contact your Logiquip authorized service representative